PROCESS TRACE OXYGEN ANALYZER CI-PC96V
CI-PC96V
CHANG AI
12 MONTHS
Vui lòng gọi
CI-PC96V Process Trace Oxygen Analyzer
Measure Range: 0-10/100/1000ppm/1% O2 FS ranges;
Sensor Type: Electrochemical fuel cell type;
Accuracy: <±1% of FS range under constant conditions;
Application: Oxygen analysis from 0.5ppm to 1% in inert, helium, hydrogen, mixed and acid
(CO2) gas streams;
Approvals: ISO9001:2008;
Area Classification: General purpose
Alarms: 2 adjustable form C relay contacts non-latching;
Calibration: Certified gas of O2 balance N2 approximating 80% of range above analysis
range recommended for optinum results; Two points calibration;
Compensation: Barometric pressure and temperature; optional heated sample system and
sensor housing;
Connections: 1/4'' or 1/8'' lock type tube fittings;
Controls: Menu driven calibration, alarm and system functions;
Data Acquisition: Selectable data point intervals;
Display: Graphical LCD 128×64MM; resolution 0.01ppm, display pressure, measure
range, real time and date;
Enclosure: 144(H)×144(W)×251(D)MM;
Flow Sensitivity: 1~1.5L/min recommended;
Pressure: Inlet - regulate to 0.5~2 bar, Vent-atmospheric;
Power: Universal: AC110~264V AC, 50/60Hz;
Response Time: 90% of final FS reading < 10 seconds;
Sample System: Flow control and sample/bypass valves; flow indicator;
Sensitivity: <0.5% of FS range;
Sensor Life: 24 months at 25℃ and 1 atm; average O2 < 1000ppm;
Signal Output: 4~20mA isolated or 0-10V (optional), etc.;
Temp. Range: 0~50℃;
Warranty: 12 months analyzer; 12 months sensor.
-
Fast recovery Insta Trace Micro-fuel Cell sensor;
-
Screen touch sensing type button operation, unique operation experience;
-
Signal output 4-20mA or 0-10V (optional), etc.;
-
Two fully-adjustable concentration alarm points with programmable relay function;
-
Full duplex RS232 communication port;
-
Universal power supply 85-265V AC 50-60Hz;
-
Extended-life, maintenance-free sensor;
-
ISO 9001 QA System.
-
Air separation and liquefaction;
-
Pure, gaseous hydrocarbon stream monitoring;
-
Semiconductor manufacturing;
-
Protective atmosphere blanketing of primary liquid feedstocks and flammable liquids;
-
Process monitoring of gaseous monomers-vinyl chloride, propylene, butadiene, isoprene or ethylene;
-
Gas purity certification.